Medical Gas Sources

CMS Medical Gas Manifold

CMS Medical Gas Manifold

A manifold is the cylinder-based gas source in a healthcare facility, used as the primary source for lower-consumption gases or as the reserve for a liquid oxygen tank or generator. CMS produces a manifold family with different operating configurations.

Overview

What is it?

The manifold comprises a duty bank and a reserve bank, cylinder tail connections, regulators reducing to pipeline pressure, non-return and over-pressure protection, and status signals wired to the alarm panel. Cylinder connections are available to different standards — British Standard, CGA or bullnose — depending on the cylinders used on site.

Features

  • Available configurations: automatic manifold, manual manifold, emergency standby manifold (ESM) and a high-flow version
  • Duty and reserve banks for continuity of supply
  • Cylinder connections to British Standard, CGA or bullnose
  • Pressure regulation, non-return valves and over-pressure protection
  • Status signals for connection to the medical gas alarm panel
  • Products from this brand are available according to project requirements and availability.

Applications

  • Hospitals and medical complexes
  • Operating theatres and intensive care units
  • Emergency and inpatient departments
  • Day-surgery centres and clinics
  • Laboratories and supporting medical applications
  • Expansion and medical gas pipeline refurbishment projects
Specifications

Technical specifications

Technical specifications and brands are set per project and per the approved specification — contact us for the technical data suited to your case.

Brand

Brand / Manufacturer

CMS products

Brand
CMS
Knowledge

Related technical articles

Medical Oxygen Systems

In preparation

Medical Gas Manifolds

In preparation

Medical Gas Testing

In preparation

FAQ

Frequently asked questions

What is the purpose of an emergency standby manifold (ESM)?

It serves as a standby source brought into use if the primary source is interrupted or during maintenance, connected to the pipeline through a dedicated connection point per the approved design.